ISOLOC 6.5 Software Part Number NT-NW-425-520. Software for radiotherapy treatment
Northwest Medical Physics Equipment Inc began a recall of ISOLOC 6.5 Software Part Number NT-NW-425-520. Software for radiotherapy treatment on . The reason FDA records is "This correction was for several software bugs which caused image format problems; incorrect ruler display; software crashes; failure to update pixel spacing; display of unattainable treatment table positions; failure to...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1146-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Northwest Medical Physics Equipment, Inc.
- Product
- ISOLOC 6.5 Software Part Number NT-NW-425-520. Software for radiotherapy treatment.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- This correction was for several software bugs which caused image format problems; incorrect ruler display; software crashes; failure to update pixel spacing; display of unattainable treatment table positions; failure to update gantry angles; wrong shifts for two of three axes of treatment table movement.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Software to fix the problems was written and software users were notified of the availability by e-mail on 04/07/2006. They could download the fixed software from the firm's internet website.
- Quantity in commerce
- 141 copies
- Distribution
- Customers in USA, Korea, Spain and UK.
- Product codes and lots
- ISOLOC 6.5 version, Part Number NT-NW-425-520