ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software

Northwest Medical Physics Equipment, Inc. / initiated 2005-05-31 / Terminated / root cause: software

Northwest Medical Physics Equipment Inc began a recall of ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software on . The reason FDA records is "When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1147-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Northwest Medical Physics Equipment, Inc.
Product
ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display values ten times greater than the correct values.
Root cause tag
software
FDA root cause
Software design
Action
Software to fix the described problems was written and customers contacted by e-mail on 05/31/2005, that a software patch, titled 05312005 was available on the firm's internet website.
Quantity in commerce
85
Distribution
Worldwide, including USA, Korea, Spain, and UK.