ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software
Northwest Medical Physics Equipment Inc began a recall of ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software on . The reason FDA records is "When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1147-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Northwest Medical Physics Equipment, Inc.
- Product
- ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display values ten times greater than the correct values.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Software to fix the described problems was written and customers contacted by e-mail on 05/31/2005, that a software patch, titled 05312005 was available on the firm's internet website.
- Quantity in commerce
- 85
- Distribution
- Worldwide, including USA, Korea, Spain, and UK.
- Product codes and lots
- ISOLOC Version 4.2 Part Number NT-NW-425-520.