ISOLOC Software Versions 6.0 and 4.2. Part Number NT-NW-425-520. Image guided radiotherapy treatment software
Northwest Medical Physics Equipment Inc began a recall of ISOLOC Software Versions 6.0 and 4.2. Part Number NT-NW-425-520. Image guided radiotherapy treatment software on . The reason FDA records is "When ISOLOC version 6.0 and 4.2 distance unit is set to centimeter, the localization coordinates were rounded to the nearest millimeter which could affect positioning by 0.5 mm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1152-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Northwest Medical Physics Equipment, Inc.
- Product
- ISOLOC Software Versions 6.0 and 4.2. Part Number NT-NW-425-520. Image guided radiotherapy treatment software.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- When ISOLOC version 6.0 and 4.2 distance unit is set to centimeter, the localization coordinates were rounded to the nearest millimeter which could affect positioning by 0.5 mm.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Software to fix the described problems was written and customers contacted by e-mail on 04/28/2005, that a software patch, titled 04282005, was available on the firm's internet website.
- Quantity in commerce
- 84
- Distribution
- Worldwide, including USA, Korea, Spain, and UK.
- Product codes and lots
- ISOLOC Software Versions 6.0 & 4.2 Part Number NT-NW-425-520
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.