Z-1152-2007

ISOLOC Software Versions 6.0 and 4.2. Part Number NT-NW-425-520. Image guided radiotherapy treatment software

Northwest Medical Physics Equipment, Inc. / initiated 2005-04-28 / Terminated / root cause: software

Northwest Medical Physics Equipment Inc began a recall of ISOLOC Software Versions 6.0 and 4.2. Part Number NT-NW-425-520. Image guided radiotherapy treatment software on . The reason FDA records is "When ISOLOC version 6.0 and 4.2 distance unit is set to centimeter, the localization coordinates were rounded to the nearest millimeter which could affect positioning by 0.5 mm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1152-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Northwest Medical Physics Equipment, Inc.
Product
ISOLOC Software Versions 6.0 and 4.2. Part Number NT-NW-425-520. Image guided radiotherapy treatment software.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
When ISOLOC version 6.0 and 4.2 distance unit is set to centimeter, the localization coordinates were rounded to the nearest millimeter which could affect positioning by 0.5 mm.
Root cause tag
software
FDA root cause
Software design
Action
Software to fix the described problems was written and customers contacted by e-mail on 04/28/2005, that a software patch, titled 04282005, was available on the firm's internet website.
Quantity in commerce
84
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Product codes and lots
ISOLOC Software Versions 6.0 & 4.2 Part Number NT-NW-425-520