ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position...
Northwest Medical Physics Equipment Inc began a recall of ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position patients) for high accuracy radiotherapy on . The reason FDA records is "When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1153-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Northwest Medical Physics Equipment, Inc.
- Product
- ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position patients) for high accuracy radiotherapy.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Software to fix the described problem was written and customers contacted by e-mail on 04/11/2005, that a software patch, titled 04112005, was available to their customers on the firm's internet website.
- Quantity in commerce
- 84
- Distribution
- Worldwide, including USA, Korea, Spain, and UK.
- Product codes and lots
- ISOLOC Software Version 6.0, Part Number NT-NW-425-520