Z-1153-2007

ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position...

Northwest Medical Physics Equipment, Inc. / initiated 2005-04-11 / Terminated / root cause: software

Northwest Medical Physics Equipment Inc began a recall of ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position patients) for high accuracy radiotherapy on . The reason FDA records is "When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1153-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Northwest Medical Physics Equipment, Inc.
Product
ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position patients) for high accuracy radiotherapy.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm.
Root cause tag
software
FDA root cause
Software design
Action
Software to fix the described problem was written and customers contacted by e-mail on 04/11/2005, that a software patch, titled 04112005, was available to their customers on the firm's internet website.
Quantity in commerce
84
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Product codes and lots
ISOLOC Software Version 6.0, Part Number NT-NW-425-520