Z-1154-2007

ISOLOC Software Version 6.0, Part Number NT-NW-425-520. Localization software for radiotherapy treatment...

Northwest Medical Physics Equipment, Inc. / initiated 2004-12-30 / Terminated / root cause: software

Northwest Medical Physics Equipment Inc began a recall of ISOLOC Software Version 6.0, Part Number NT-NW-425-520. Localization software for radiotherapy treatment system on . The reason FDA records is "When ISOLOC version 6.0 software was used in the Anatomic Landmark mode, the couch moves display in the Couch Translator window was incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1154-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Northwest Medical Physics Equipment, Inc.
Product
ISOLOC Software Version 6.0, Part Number NT-NW-425-520. Localization software for radiotherapy treatment system.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
When ISOLOC version 6.0 software was used in the Anatomic Landmark mode, the couch moves display in the Couch Translator window was incorrect.
Root cause tag
software
FDA root cause
Software design
Action
Software to fix the described problem was written and customers contacted by e-mail on 12/30/2004 that a software patch, titled 12302004 was available to their customers on the firm's internet website.
Quantity in commerce
79
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Product codes and lots
ISOLOC Software Version 6.0, Part Number NT-NW-425-520