Z-1155-2007

ISOLOC Software Version 4.2, Part Number NT-NW-425-520. Image guided radiation treatment software

Northwest Medical Physics Equipment, Inc. / initiated 2004-11-30 / Terminated / root cause: software

Northwest Medical Physics Equipment Inc began a recall of ISOLOC Software Version 4.2, Part Number NT-NW-425-520. Image guided radiation treatment software on . The reason FDA records is "When ISOLOC Version 4.2 unit is set to centimeter and the user clicks the Reenter button in the digitizer window, the coordinates of the markers and target in the anatomic imaging system erroneously displays values ten...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1155-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Northwest Medical Physics Equipment, Inc.
Product
ISOLOC Software Version 4.2, Part Number NT-NW-425-520. Image guided radiation treatment software.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
When ISOLOC Version 4.2 unit is set to centimeter and the user clicks the Reenter button in the digitizer window, the coordinates of the markers and target in the anatomic imaging system erroneously displays values ten times greater than the correct values.
Root cause tag
software
FDA root cause
Software design
Action
Software to fix the described problem was written and customers contacted by e-mail on 11/30/2004, that a software patch, titled 11302004, was available to their customers on the firm's internet website.
Quantity in commerce
83
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Product codes and lots
ISOLOC Software Version 4.2, Part Number NT-NW-425-520.