Z-0002-2008
FP1000 Cell Preparation System Software versions 1.1 & 1.2, Part Number 624922, for in vitro diagnostic use
Siemens Healthcare Diagnostics Inc. / initiated 2007-02-22 / Terminated / root cause: labeling and UDI
Beckman Coulter Inc began a recall of FP1000 Cell Preparation System Software versions 1.1 & 1.2, Part Number 624922, for in vitro diagnostic use on . The reason FDA records is "Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0002-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- FP1000 Cell Preparation System Software versions 1.1 & 1.2, Part Number 624922, for in vitro diagnostic use.
- Product code
- LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use
- Reason for recall
- Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- A Product Corrective Action (PCA) letter was mailed on Feb 22, 2007, to all FP1000 customers to inform them of an issue associated with the FP1000 Cell Preparation System. Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented. Under these conditions the FP1000 will under-dispense specimen volume leading to lower than expected absolute counts. Specimen tubes that are recapped or underfilled must therefore be vented prior to running on the FP1000. If the initial state of the specimen tube is unknown, then it must be vented to ensure accurate results prior to running on the FP1000. Additional instructions are provided. A Fax Back Response Form was also provided.
- Quantity in commerce
- 9 in the U.S.; 1 in Canada
- Distribution
- Nationwide (MA, IN, IL, UT, RI, NJ), and Canada.
- Product codes and lots
- Part Number 624922. Software versions 1.1 & 1.2