Z-1174-2007
EasyLink Informatics System, software version 3.0, Part Number 1000034941, DADE BEHRING INC., Newark, DE
Siemens Healthcare Diagnostics Inc. / initiated 2007-07-03 / Terminated
Dade Behring, Inc. began a recall of EasyLink Informatics System, software version 3.0, Part Number 1000034941, DADE BEHRING INC., Newark, DE on . The reason FDA records is "Sample ID and Patient results could be associated with the wrong patient". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1174-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- EasyLink Informatics System, software version 3.0, Part Number 1000034941, DADE BEHRING INC., Newark, DE.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Sample ID and Patient results could be associated with the wrong patient
- FDA root cause
- Other
- Action
- The recalling firm issued a Field Corrective Action letter dated 7/3/07 to inform their customers of the problem and that a software patch would be sent for them to download. The recalling firm plans to telephone the customers after the patch has been released to verify that it has been installed.
- Quantity in commerce
- 72 units
- Distribution
- Worldwide: USA including states of AZ, CA, FL, IL, MA. MI, MO, NC, OH, PA, TX, and WV. The product was also shipped to other Dade facilities in MO, DE, IL, CT, and Belgium.
- Product codes and lots
- Part number 1000034941