Z-1174-2007

EasyLink Informatics System, software version 3.0, Part Number 1000034941, DADE BEHRING INC., Newark, DE

Siemens Healthcare Diagnostics Inc. / initiated 2007-07-03 / Terminated

Dade Behring, Inc. began a recall of EasyLink Informatics System, software version 3.0, Part Number 1000034941, DADE BEHRING INC., Newark, DE on . The reason FDA records is "Sample ID and Patient results could be associated with the wrong patient". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1174-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
EasyLink Informatics System, software version 3.0, Part Number 1000034941, DADE BEHRING INC., Newark, DE.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Sample ID and Patient results could be associated with the wrong patient
FDA root cause
Other
Action
The recalling firm issued a Field Corrective Action letter dated 7/3/07 to inform their customers of the problem and that a software patch would be sent for them to download. The recalling firm plans to telephone the customers after the patch has been released to verify that it has been installed.
Quantity in commerce
72 units
Distribution
Worldwide: USA including states of AZ, CA, FL, IL, MA. MI, MO, NC, OH, PA, TX, and WV. The product was also shipped to other Dade facilities in MO, DE, IL, CT, and Belgium.
Product codes and lots
Part number 1000034941