Z-1184-2007

Arizant Healthcare Ranger Standard Flow Blood/Fluid Warming Disposable Sets with injection port and extension...

Arizant Healthcare, Inc. / initiated 2007-06-26 / Terminated / root cause: packaging

Arizant Healthcare, Inc began a recall of Arizant Healthcare Ranger Standard Flow Blood/Fluid Warming Disposable Sets with injection port and extension set, Part Number 24250. For use with a Ranger... on . The reason FDA records is "Incomplete seal on the product pouch; with the seal incomplete, the sterility of the product inside the pouch cannot be guaranteed". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1184-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arizant Healthcare, Inc.
Product
Arizant Healthcare Ranger Standard Flow Blood/Fluid Warming Disposable Sets with injection port and extension set, Part Number 24250. For use with a Ranger warming unit.
Product code
FPA Set, Administration, Intravascular
Reason for recall
Incomplete seal on the product pouch; with the seal incomplete, the sterility of the product inside the pouch cannot be guaranteed.
Root cause tag
packaging
FDA root cause
Other
Action
All consignees will be contacted by Urgent Medical Device Recall letter on 6/26/07.
Quantity in commerce
12,000
Distribution
Worldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL., IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and countries of United Kingdom, France, Switzerland, Germany, India, Canada, Australia, Netherlands, Italy, Sweeden, South Korea, and Greece.
Product codes and lots
Lots B4798 through B4803 and Lot E5111