BMSI "MicroJack" PN 672-510000 patient interface electrode connection device used during...
Nicolet Biomedical Div of Viasys Healthcare began a recall of BMSI "MicroJack" PN 672-510000 patient interface electrode connection device used during electroencephalography (EEG) /Epilepsy Monitoring evaluations with... on . The reason FDA records is "The BMSI "MicroJack" patient interface electrode connection device used with the EEG monitor systems was found to be mis-wired. A complainant alleged that during depth electrode EEG study of a patient in preparation for...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1225-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- BMSI "MicroJack" PN 672-510000 patient interface electrode connection device used during electroencephalography (EEG) /Epilepsy Monitoring evaluations with various EEG monitoring systems. The MicroJack is used with BMSI Models 2000, 3000, 4000, 5000 & 6000, and is used with the BMSI kit upgraded with NicoletOne. Nicolet Biomedical, Inc.
- Product code
- GWQ Full-Montage Standard Electroencephalograph
- Reason for recall
- The BMSI "MicroJack" patient interface electrode connection device used with the EEG monitor systems was found to be mis-wired. A complainant alleged that during depth electrode EEG study of a patient in preparation for brain surgery there was an observation that was perceived as non-logical data output.
- FDA root cause
- Other
- Action
- VIASYS will initiate a field correction for units that have been distributed. The correction consists of testing the units and replacing those that have been found to contain the defect. Method of communication will be by certiified mail to identify consignees and receipt of the communication will be tracked. Letters "Viasys HealthCare Urgent Medical Device Field Correction (Recall)" were sent 7/17/07 to both domestic and international customers.
- Quantity in commerce
- 2102
- Distribution
- Worldwide, including USA, ISRAEL, LITHUANIA, RUSSIA, FRANCE, IRELAND, UNITED KINGDOM , SPAIN, YEMEN, JAPAN, TAIWAN, KOREA, CHINA, SAUDI ARABIA, HONG KONG, INDIA, COLOMBIA, BRAZIL, GERMANY, CANADA, DENMARK, ITALY, MEXICO, and THAILAND.
- Product codes and lots
- Models BMSI 2000, 3000, 4000, 5000 & 6000 and BMSI kit upgraded with NicoletOne.