Terumo Advanced Perfusion System, 1 Roller Pump, 4 inch diameter; Catalog No. 801040, Terumo Cardiovascular...

Terumo Cardiovascular Systems Corporation / initiated 2007-09-10 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System, 1 Roller Pump, 4 inch diameter; Catalog No. 801040, Terumo Cardiovascular Systems, Ann Arbor, MI 48103 on . The reason FDA records is "Failure to operate; The pump may fail to operate, and be difficult to hand crank, due to an internal electrical short condition". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0321-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System, 1 Roller Pump, 4 inch diameter; Catalog No. 801040, Terumo Cardiovascular Systems, Ann Arbor, MI 48103
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Failure to operate; The pump may fail to operate, and be difficult to hand crank, due to an internal electrical short condition.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified via an Urgent Medical Device Correction letter dated 9/10/07, which provided them guidance in the event the problem occurs and advised them they would be contacted to schedule replacement of the circuit boards.
Quantity in commerce
16
Distribution
Worldwide Distribution-USA including states of Alabama, Arizona, Florida, Illinois, Indiana, Mississippi, Nebraska, North Carolina, Pennsylvania, and Virginia, and countries of Belgium, Hong Kong and United Arab Emirates.
Product codes and lots
Serial Numbers 2110 through 2125.