Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog Number 801046, Terumo...
Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog Number 801046, Terumo Cardiovascular Systems, Ann Arbor, MI 48103 on . The reason FDA records is "Failure to operate: The unit display may go blank and reset when the start button is pressed, due to an internal electrical short, resulting in a centrifugal pump that fails to start as expected". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0323-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog Number 801046, Terumo Cardiovascular Systems, Ann Arbor, MI 48103
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Failure to operate: The unit display may go blank and reset when the start button is pressed, due to an internal electrical short, resulting in a centrifugal pump that fails to start as expected.
- FDA root cause
- Component change control
- Action
- Consignees were notified via an Urgent Medical Device Correction letter dated 9/10/07, given instructions to follow if the problem were to manifest itself, and informed the firm would contact them to schedule an upgrade.
- Quantity in commerce
- 120
- Distribution
- Nationwide.
- Product codes and lots
- Serial Numbers 0020 through 0041, 0043, 0044, 0046 through 0060, 0062 through 0091, 0094, 0098, 0118 through 0137, 0139 through 0156, 0159 and 0166 through 0176.