Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1240-2007

BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002...

Abbott Medical / initiated 2007-07-18 / Terminated

Smiths Medical PM, Inc. began a recall of BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part... on . The reason FDA records is "Phantom pacer spikes caused by a damaged resistor". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1240-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc
Product code
DQA Oximeter
Reason for recall
Phantom pacer spikes caused by a damaged resistor.
FDA root cause
Other
Action
Customers will be notified via a Customer Information Bulletin CIB Tracking Number: 07-CIB 12, July 18,2007 requesting them to evalute the functionality of their monitor (s). If the monitor does not exhibit any false pace detect makers when pace detect is enabled and respiration is disabled, no further action is required. If it does exhibit false pace detect they will be instructed to contact the Smiths Medical PM, Inc service department. At that time Technical Worksheet, 07-TW02, will be provided. The consignee will need to verify they are capable of perfoming the Technical Worksheet provided. The Smiths Medical PM. Inc. Service Manger will veirfy if the consignee is capable of performing the rework.
Quantity in commerce
178
Distribution
World wide: USA including states of KY, MS, PA, LA, and PR and countries of Poland, Mexico, Kuwait, Taiwan, Singapore, Hon Kong, India, Bolivia, Colombia, Chile, Argentia, Brunei Darssalam, Israel, Greece, Romania, Germany, The Netherlands, France, Great Britain, Spain, Brazil, Russia, China.
Product codes and lots
Catalog number/ serial number: 3404001/AI05110052/ AI05120000/ AI05120011/ AI05120012/ AI05120013/ AI05120018/ AI06060015/ AI06070047/ AI06100004/ AI06100005/ AI06100006/ , 3404001S/AI05110000/ AI05110001/ AI05110002/ AI05110003/ AI05110004/ AI05110005/ AI05110006/ AI05110007/ AI05110008/ AI05110009/ AI05110010/ AI05110011/ AI05110012/ AI05110013/ AI05110014/ AI05110015/ AI05110016/ AI05110017/ AI05110018/ AI05110019/ AI05110020/ AI05110021/ AI05110022/ AI05110023/ AI05110024/ AI05110025/ AI05110026/ AI05110027/ AI05110028/ AI05110029/ AI05110030/ AI05110031/ AI05110032/ AI05110033/ AI05110034/ AI05110035/ AI05110036/ AI05110037/ AI05110038/ AI05110039/ AI05120010/ AI05120016/ AI05120017/ AI05120019/ AI06010042/ AI06010043/ AI06060003/ AI06060004/ AI06060005/ AI06060006/ AI06060007/ AI06060008/ AI06060009/ AI06060010/ AI06060011/ , 3404002/AI05090021/ AI05090024/ AI05110049/ AI05120015/ AI06010035/ AI06010036/ AI06010049/ AI06040000/ AI06040001/ AI06040002/ AI06040003/ AI06040004/ AI06040005/ AI06040006/ AI06040007/ AI06040008/ AI06040009/ AI06040010/ AI06040011/ AI06040012/ AI06040013/ AI06040014/ AI06040015/ AI06040016/ AI06040017/ AI06040018/ AI06040020/ AI06060012/ AI06060013/ AI06060014/ AI06060017/ AI06060018/ AI06060019/ AI06060020/ AI06070037/ AI06070038/ AI06070039/ AI06070040/ AI06070042/ AI06070043/ AI06070044/ AI06070045/ AI06090053/ AI06100011/ AI06100012/ AI07010057/ AI07010058/ , 3404002E/AI05110040 AI05110041/ AI05110042/ AI05120001/ AI05120002/ AI05120005/ AI05120006/ AI05120007/ AI05120008/ AI05120009/ AI05120024/ AI06010041/ AI06010044/ AI06010045/ AI06010046/ AI06010048/ AI06040019/ AI06040023/ AI06040024/ AI06070041/ AI06070046/ AI06070048/ AI06070049/ AI06090050/ AI06090051/ , 3404002F/AI07010067/ AI07010068/ , 3404002G/AI05110050/ AI05110051/ AI05120003/ AI05120004/ AI06060021/ AI06060022/ AI06060023/ AI06070035/ AI06070036/ AI06090054/ AI06090055/ AI06090056/ AI06090057/ AI06090058/ AI06100008/ AI06100009/ AI06100010/ AI06100018/ , 3404002S/AI05110043/ AI05110044/ AI05110045/ AI06010037/ AI06010038/ AI06010039/ AI06010040/ AI06060000/ AI06060001/ AI06060002/ AI06060016/ AI06090052/ AI06100019/ AI06100020/ AI06100021/ AI06100022/ AI06100023/ AI06100024/ AI07010050/ AI07010051/