Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1227-2007

Medtronic Intersept 3608 Sterile tubing (8') 1/2 x 3/32 in. Quantity 10, Lot 7626621, Use by 2009-07-31...

Medtronic Perfusion Systems / initiated 2007-08-01 / Terminated

Medtronic Perfusion Systems began a recall of Medtronic Intersept 3608 Sterile tubing (8') 1/2 x 3/32 in. Quantity 10, Lot 7626621, Use by . Manufacturer: Medtronic, Inc. Minneapolis, MN... on . The reason FDA records is "Medtronic Custom Tubing Packs might have open header bag seals". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1227-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Perfusion Systems
Product
Medtronic Intersept 3608 Sterile tubing (8') 1/2 x 3/32 in. Quantity 10, Lot 7626621, Use by 2009-07-31. Manufacturer: Medtronic, Inc. Minneapolis, MN 55432-5604 USA. Different components and tubing sets are included within the Custom Pack depending on customer requirements. Tubing sets are placed within a plastic tray, which then is wrapped and placed inside a header bag and sealed at the plastic side. This bag seves as the product sterile barrier.
Product code
DWE Tubing, Pump, Cardiopulmonary Bypass
Reason for recall
Medtronic Custom Tubing Packs might have open header bag seals.
FDA root cause
Other
Action
Certified letters dated 8/1/07 were sent to customers. The letter explains the situation and requests the impacted product to be returned to Medtronic for replacement or credit. The customers are being requested to complete a Recall Certificate acknowledging the receipt of the information.
Quantity in commerce
853 packs
Distribution
Nationwide.
Product codes and lots
Custom packs with catalogue# / lot #:  3608/7626621, CB2E25R7/7760483, CB2E28R/7760491, CB4Q86R/7226272, 9156R2/8105907, 1D41R10/7372026, 1F61R4 / 7038360, 1E04R7 / 7277651, 9156R2 / 8105907, 1X37R / 7760774, 3608 / 7626621, 0X01R7 / 7278478,  2567R41/ 6577695, 1E97R2 / 7828564, 5G01R / 7734883, CB5305R1 / 7346602, CB4Q86R / 7226272,  2T10R1 / 7252235, 1D91R1 / 7843300, 1A45R3 /7584969, 3M16R5 / 7411494, CB2J35R3 / 7352341,  CB3R62R / 7668062, 4V60R / 7934463, DLP2X90R /7504957, 3608 /7626621, 3608/ 7626621, 3C86R9/ 7656387,  3608/ 7626621, 3608/ 7626621, HY2K51R14/ 7438662, 2H08R2/ 7668273, 3608/ 7626621, 1A05R15/ 7038597, CB4826R3 / 7843481, 3608/ 7626621, TL4K84R7/ 7900108, CB3K45R6/ 7626276, 9295R2/ 7934447, 2100R3/ 7694869, CB3U51R7 / 7695001, 9156R2/ 8105907, 9182R3/ 7694914, 0M14R12/ 7694762, 2C50R/ 7760791, 3608/ 7626621, 2K98R1/ 7668290, CB0U83R11/ 7721206,  2997R27/ 7438443, 9156R2/ 8105907, TL3E97R5/ 7722225, TL4J32R8/ 7761048, 2Z75R5/ 7960821, CB2E25R7 /7760483, 4W91R/ 7843431, 3608/ 7626621, 9156R2/ 8105907, 1G69R7/ 7760715, 3608/ 7626621, 2T28R1/ 7899872, 9156R2/ 8105907, 1F77R3/ 7722348, 1F77R3/ 7722348, 1F77R3/ 7722348, DLP4X54R/ 7761021, 3608/ 7626621, 3608/ 7626621