Tosoh Bioscience Inc began a recall of 4 glass vials per box. 2/4ml vials of Sample Diluting Solution & 2/2ml vials of Calibration Verification Material. BetaHCG Calibration Verification Test Set... on . The reason FDA records is "Firm discovered that a deterioration value of approximately 20% was found when the product of this lot was measured. Tosoh AIA confirmed that deterioration of the product occurred over time and resulted in a deviation...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0007-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tosoh Bioscience Inc
- Product
- 4 glass vials per box. 2/4ml vials of Sample Diluting Solution & 2/2ml vials of Calibration Verification Material. BetaHCG Calibration Verification Test Set, in vitro diagnostic.
- Product code
- JJX Single (Specified) Analyte Controls (Assayed And Unassayed)
- Reason for recall
- Firm discovered that a deterioration value of approximately 20% was found when the product of this lot was measured. Tosoh AIA confirmed that deterioration of the product occurred over time and resulted in a deviation from the assigned value.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Letter, dated July 2, 2006. Immediately examine inventory and quarantine product. If product was further distributed, the direct accounts are to notify their customers including a copy of the original recall letter and Notification of Recall response card.
- Quantity in commerce
- 47 Boxes
- Distribution
- Nationwide.
- Product codes and lots
- Catalog # 020661 Lot #G803536 Exp. Date August 2008