Z-1229-2007

Guidewire, 70cm x 1mm, REF 5583705, Sterile, Single Use, BARD Access Systems, Hemodialysis Catheter, Salt...

Bard Access Systems, Inc. / initiated 2007-08-10 / Terminated

C.R. Bard Access Systems, Inc began a recall of Guidewire, 70cm x 1mm, REF 5583705, Sterile, Single Use, BARD Access Systems, Hemodialysis Catheter, Salt Lake City, UT 84116 on . The reason FDA records is "Metal guidewires used in Hemodialysis Catheters may have been reverse loaded within the dispenser. Potential exists for the stiff end of the guidewire to be inserted into the patient instead of the flexible end". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1229-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Access Systems, Inc.
Product
Guidewire, 70cm x 1mm, REF 5583705, Sterile, Single Use, BARD Access Systems, Hemodialysis Catheter, Salt Lake City, UT 84116.
Product code
MPB Catheter, Hemodialysis, Non-Implanted
Reason for recall
Metal guidewires used in Hemodialysis Catheters may have been reverse loaded within the dispenser. Potential exists for the stiff end of the guidewire to be inserted into the patient instead of the flexible end.
FDA root cause
Other
Action
Consignees were notified by letter on 08/10/2007 and asked to remove and return all unused affected product to Bard.
Quantity in commerce
7225 units
Distribution
Worldwide, including USA, Canada, Latin America, and Europe (Austria, Belgium, Finland, France, Germany, Great Britain, Holland, Italy, Norway, Spain, and Sweden).
Product codes and lots
Product Code 5583705, Lot Numbers: 22FMA202, 22IMO531, 22KMO573, 22DN2914, 22FN5232, 22GN5835, 22GN5878, 22HN6574, 22BO6078, 22BO6163, 22FO6086, 22GO6026, 22HO6101, 22IO6111, 22KO6124, 22CP6268, 22EP6831, REPG0035, REPG0643, REPH0954, REPK0058, REPK0729, REPL0298, REQB0030, REQB0224, REQD0200, REQF0429, REQF0736, REQJ0282, REQK0875, RERC0052, RERD0719.