Z-1206-2007

Spectrum Designs Medical Anatomical I (Silhouette) Malar Facial Implant, Catalog Number 120-224S

Spectrum Designs, Inc. / initiated 2007-06-15 / Terminated / root cause: packaging

Spectrum Designs Inc began a recall of Spectrum Designs Medical Anatomical I (Silhouette) Malar Facial Implant, Catalog Number 120-224S on . The reason FDA records is "The product is being recalled because it was discovered the at least one unit in the lot contained a mismatched pair (two rights and no left). Since the lot was produced with equal numbers of pairs, it is assumed that...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1206-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Spectrum Designs, Inc.
Product
Spectrum Designs Medical Anatomical I (Silhouette) Malar Facial Implant, Catalog Number 120-224S.
Product code
LZK Implant, Malar
Reason for recall
The product is being recalled because it was discovered the at least one unit in the lot contained a mismatched pair (two rights and no left). Since the lot was produced with equal numbers of pairs, it is assumed that one other unit was packaged with two lefts and no right. This was discovered through a physician complaint.
Root cause tag
packaging
FDA root cause
Other
Action
The firm initiated the recall on 06/15/2007 by contacting their customers by telephone. The firm followed-up by sending recall letters via FedEx or certified mail with signature confirmation to each direct account who received the recalled product on 06/18/2007.
Quantity in commerce
15 pairs (left/right)
Distribution
Nationwide.
Product codes and lots
Lot number 30838, serial numbers BAE256 through BAE275, manufactured June, 2006. Expiration date is five years from date of manufacture.