Advanced Medical Optics (AMO), VISX WaveScan WaveFront System (software). Part Number 0070-1478. Advanced...
Visx Inc began a recall of Advanced Medical Optics (AMO), VISX WaveScan WaveFront System (software). Part Number 0070-1478. Advanced Medical Optics, Santa Clara, CA 95051 on . The reason FDA records is "Erroneous Treatment Calculations - Two software caused errors; in combination, in the WaveScan System Software, will result in an erroneous treatment calculation (overcorrection) in patients. (1) Installation of...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0365-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Visx Incorporated, a Subsidiary of AMO Inc
- Product
- Advanced Medical Optics (AMO), VISX WaveScan WaveFront System (software). Part Number 0070-1478. Advanced Medical Optics, Santa Clara, CA 95051.
- Product code
- LZS Excimer Laser System
- Reason for recall
- Erroneous Treatment Calculations - Two software caused errors; in combination, in the WaveScan System Software, will result in an erroneous treatment calculation (overcorrection) in patients. (1) Installation of software for a certain brand of computer (Dedicated Computing LLC) can cause a software registry setting for an algorithm that identifies image reflection to be erroneously set in the "off
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- AMO sent an Action Required notification on May 30, 2007, to AMO Certified Physicians. The letter included a CD software update (0070-1721) to be installed by users that provided correct registry parameter settings. Upon completion of installation, a confirmation form will print out. Customers are requested to fax back confirmation to AMO (408-773-7055).
- Quantity in commerce
- 196 systems are affected in the U.S. and 133 outside of the U.S.
- Distribution
- Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO,MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK,OR,PA, SC, TN, TX, UT, VA, WA, WI, and WV and countries of Canada and Mexico.
- Product codes and lots
- System Serial Numbers affected include: 75447, 75448, 76086, 75389, 70170, 76092, 76004, 76020, 76070, 76128, 75391, 75413, 75416, 75417, 75423, 75441, 75442, 76043, 76046, 76048, 76082, 76091, 76121, 76144, 76147, 75419, 76122, 75257, 75385, 75386, 76003, 76022, 76111, 76018, 75390, 76050, 76098, 75420, 76071, 76075, 76114, 75411, 75425, 76123, 76124, 76136, 75384, 75421, 76002, 76023, 75449, 76084, 76119, 76085, 76033, 76036, 76083, 76076, 76052, 76089, 76067, 75444, 75450, 76047, 76099, 76146, 76026, 75424, 76095, 75410, 76106, 75392, 75409, 75428, 75452, 76025, 76032, 76045, 76053, 76097, 70213, 70256, 70316, 70776, 75358, 75418, 75430, 75431, 75432, 76042, 76044, 76049, 76051, 76077, 76078, 76108, 76109, 76116, 76126, 76141, 76035, 76087, 76112, 75398, 76115, 76013, 76120, 76057, 75387, 76021, 76125, 76127, 76130, 76135, 75422, 76069, 75415, 76110, 76137, 76161, 76163, 70419, 70869, 75403, 76096, 76142, 70304, 70160, 70250, 70666, 70753, 70890, 75028, 75402, 75405, 75426, 75429, 75436, 76007, 76027, 76029, 76031, 76037, 76065, 76068, 76073, 78074, 76079, 76103, 76117, 76139, 76140, 76143, 76150, 76151, 76153, 76155, 76156, 76157, 76159, 76160,76162, 70272, 70280, 70297, 70706, 70872, 75297, 75359, 75404, 76059, 76061, 76063, 76081, 76149, 75397, 75399, 70541, 70774, 70904, 75412, 75437, 76017, 70043, 70394, 75401, 75439, 76066, 76080, 76088, 76148, 76058, 76090, 70271, 70353, 70398, 70409, 70734, 76006, 76009, 76028, 76154, 76158, 76145, 70855, 76038, 76062, 75433, 70864, 75394, 75434, 76138, 71704, 70217, 70485, 70491, 70690, 75281, 76104, 75406, 75414, 75427, 76015, 70286, 70440, 70476, 70765, 70882, 75400, 75446, 76005, 75388, 76024, 76039, 76129, 70588, 70603, 75445, 76040, 70597, 70175, 70178, 76094, 70759, 70336, 70637, 75230, 76093, 76105, 75121, 76019, 76056, 76060, 70333, 70341, 70365, 70379, 70428, 75113, 76001, 76054, 70691, 75109, 76107, 70299, 75234, 70554, 70653, 75236, 76011, 76152, 70474, 70495, 70542, 75440, 76101, 76118, 76113, 75435, 76012, 70031, 70032, 70036, 70092, 70227, 70580, 70608, 70721, 75172, 75395, 75408, 75443, 75451, 76016, 76030, 76034, 76072, 76100, 76064, 70080, 70221, 70283, 75328, 75396, 75438, 75407, 70159, 70241, 70350, 76008, 76010, 76014, 70650, 70732, 76055, 76102, 70457, and 76041