Z-0010-2008
EPIK KNEE SYSTEM Angled Insert Impactor RM/LL; Part #801-03-052 Rev. A; non-sterile; knee prosthesis...
Encore Medical, LP / initiated 2007-08-09 / Terminated / root cause: design
Encore Medical, Lp began a recall of EPIK KNEE SYSTEM Angled Insert Impactor RM/LL; Part #801-03-052 Rev. A; non-sterile; knee prosthesis instrument. Distributed by Encore Medical, L.P., Austin... on . The reason FDA records is "Angled Insert Impactor may scratch Tibial Insert on the articulating surface, causing increased wear at the damaged site". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0010-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- EPIK KNEE SYSTEM Angled Insert Impactor RM/LL; Part #801-03-052 Rev. A; non-sterile; knee prosthesis instrument. Distributed by Encore Medical, L.P., Austin, TX 78758.
- Product code
- HWA Impactor
- Reason for recall
- Angled Insert Impactor may scratch Tibial Insert on the articulating surface, causing increased wear at the damaged site.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Firm notified consignees regarding recall via letter on 08/09/07. Consignees informed that Encore customer service representatives would be contacting them to exchange recalled units with replacements.
- Quantity in commerce
- 5 units.
- Distribution
- Products distributed to direct accounts in CA, FL and MD.
- Product codes and lots
- All lots.