Z-0014-2008
Brain Heart Infusion Agar labeled BHIA + Vancomycin 6 MCG/ML For Screening for Vancomycin Resistant...
PML, Inc. dba Pml Microbiologicals Inc. / initiated 2007-04-23 / Terminated / root cause: manufacturing process
Pml Microbiologicals Inc. began a recall of Brain Heart Infusion Agar labeled BHIA + Vancomycin 6 MCG/ML For Screening for Vancomycin Resistant Enterococci PML microbiologicals Wilsonville, Oregon 97070... on . The reason FDA records is "Plates manufactured without vancomycin. Possibility for false resistant results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0014-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- PML, Inc. dba Pml Microbiologicals Inc.
- Product
- Brain Heart Infusion Agar labeled BHIA + Vancomycin 6 MCG/ML For Screening for Vancomycin Resistant Enterococci PML microbiologicals Wilsonville, Oregon 97070 USA Catalog #P1174
- Product code
- JSO Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar
- Reason for recall
- Plates manufactured without vancomycin. Possibility for false resistant results.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- On 4/23/07 the firm began contacting customers via fax and telephone.
- Quantity in commerce
- 640 total plates, 320 per lot (lot 229957-1 was dist only in Canada; lot 229958-1 was dist in the US and Canada)
- Distribution
- lot #229957-1: 120 plates were distributed to 9 Canadian consignees; 200 plates were shipped, quarantined, and destroyed at the Canadian distribution center. lot #229958-1: 110 plates were distributed to 9 US consignees; 210 plates were shipped, quarantined, and destroyed at the Canadian distribution center.
- Product codes and lots
- Lot numbers 229957-1 and 229958-1