Z-0015-2008
Ammonia Liquid Stable Reagent, Enzymatic Method. Catalog number TR60101. in vitro diagnostic
Thermo Fisher Scientific Oy / initiated 2007-07-24 / Terminated / root cause: design
Thermo Fisher Scientific began a recall of Ammonia Liquid Stable Reagent, Enzymatic Method. Catalog number TR60101. in vitro diagnostic on . The reason FDA records is "Method is susceptible to positive interference from endogenous ALT in patient samples". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0015-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Thermo Fisher Scientific Oy
- Product
- Ammonia Liquid Stable Reagent, Enzymatic Method. Catalog number TR60101. in vitro diagnostic.
- Product code
- JIF Enzymatic Method, Ammonia
- Reason for recall
- Method is susceptible to positive interference from endogenous ALT in patient samples.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm has updated the Limitations section of the Ammonia Reagent Package Insert and circulated a Technical Bulletin dated July 24, 2007 to all customers that have purchased Ammonia Reagent that is still within expiration date. A Notification of Change has been introduced and shall be placed within subsequent Ammonia Reagent Kits with a copy of the updated package insert; the notification of change informs customers that a change has been made to the package insert. Notifications of Change are placed within kits for 6 months from the date of issue.
- Quantity in commerce
- 6,882 units
- Distribution
- Worldwide, including USA, Ecuador, and Canada.
- Product codes and lots
- Lot numbers FL029 exp 8/30/07, GB059 exp 11/30/07, GD005 exp 12/30/07, GE093 exp 1/30/08, GH034 exp 4/30/08, GJ039 exp 7/30/08, HB064 exp 11/30/08, and HD035 exp 1/30/09.