Z-0056-2008

Angiomat Illumena Injector with Auxiliary Console, Mallinckrodt, Inc., Hazelwood, MO

Mallinckrodt, Inc. / initiated 2007-08-24 / Terminated / root cause: design

Mallinckrodt Inc began a recall of Angiomat Illumena Injector with Auxiliary Console, Mallinckrodt, Inc., Hazelwood, MO on . The reason FDA records is "Synchronization problems: units that have both a main programmable console and an auxiliary programmable console, when simultaneously, injection parameters are programmed into the unit's main console and the unit's...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0056-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mallinckrodt, Inc.
Product
Angiomat Illumena Injector with Auxiliary Console, Mallinckrodt, Inc., Hazelwood, MO.
Product code
DXT Injector And Syringe, Angiographic
Reason for recall
Synchronization problems: units that have both a main programmable console and an auxiliary programmable console, when simultaneously, injection parameters are programmed into the unit's main console and the unit's auxiliary consoles, mis-synchronization could occur resulting in a variance between the intended injection parameters (those programmed) and the actual parameters of the injection deliv
Root cause tag
design
FDA root cause
Device Design
Action
On 9/20/2007, a notification was sent from Mallinckrodt to all direct accounts via certified mail. The notification details the potential for mis-synchronization of injection parameters if the main and auxiliary consoles are simultaneously programmed. The notification reminds end-users that simultaneous programming should not be performed and to ensure it is not performed, a notification label (provided in the notification) is to be affixed on both consoles of the injector system. Once the label is affixed to the system, end-users are instructed to complete the notification survey and return the survey to the Mallinckrodt. In addition, the notification includes an updated chapter insert (Chapter 11.1) for the injector system's product manual that emphasize the separation between programming of the main and auxiliary consoles. Questions regarding this matter may be directed to Mallinckrodt's Product Monitoring Department at 1-888-744-1414, menu option 1 and then menu option 2.
Quantity in commerce
176
Distribution
Worldwide to include:U.S.A. states of - AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, PA, TN, TX, UT, VA, WA, & WI. and Foreign countries of Argentina, Belgium, Brazil, Canada, Germany, Japan, Singapore, & Zambia.
Product codes and lots
Product Model Numbers : 902040; 904045; and 902045CI