Z-1224-2007
Baxter Colleague CX Single Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication...
Baxter Healthcare Corporation / initiated 2007-08-15 / Terminated
Baxter Healthcare Corp. began a recall of Baxter Colleague CX Single Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in... on . The reason FDA records is "Some of the pumps were upgraded by one service technician who failed to replace prisms and batteries as required in the upgrade". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1224-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Colleague CX Single Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore; product code 2M8161
- Product code
- FRN Pump, Infusion
- Reason for recall
- Some of the pumps were upgraded by one service technician who failed to replace prisms and batteries as required in the upgrade.
- FDA root cause
- Other
- Action
- The affected hospitals were sent an Urgent Colleague Deployment Correction letter dated 8/15/07 via overnight mail. The hospitals were informed that some of the Colleague pumps at their facility may not have been completely upgraded to include all the necessary components required as part of Baxter's Corrective Action Plan (CAP). The hospitals were provided a list of affected pump serial numbers and were requested to remove the pumps from clinical use and notify all end users of the action. The accounts were requested to complete the enclosed customer reply form, acknowledging the receipt and understanding of the letter and fax it back to Baxter at 1-847-270-5457. The hospitals were also requested to report any malfunction or adverse event related to the Colleague Infusion Pumps to Baxter Corporate Product Surveillance at 1-800-437-5176.
- Quantity in commerce
- 42 pumps
- Distribution
- Florida and California
- Product codes and lots
- product code 2M8161, serial numbers 14051565CC, 15034273CC, 15034492CC, 15040255CC, 15041116CC, 15041169CC, 15041170CC, 15041172CC, 15041181CC, 15041194CC, 15041199CC, 15041206CC, 15041212CC, 15041218CC, 15041241CC, 15041288CC, 15041290CC, 15041317CC, 15050020CC, 15050034CC, 15050045CC, 15050061CC, 15050080CC, 15050117CC, 15050129CC, 15060087CC, 15060217CC, 15060253CC, 15060479CC, 15060489CC, 15060549CC, 15060557CC, 15060600CC, 15060738CC, 15060790CC, 15060827CC, 15060939CC, 15060940CC, 15060968CC, 15063211CC, 15075371CC, 15075103CC