Z-1212-2007
The Pinnacle3 Radiation Therapy Planning (RTP) system and its associated software; composed of several...
Philips North America LLC / initiated 2007-08-21 / Terminated / root cause: software
Philips Nuclear Medicine, Inc. began a recall of The Pinnacle3 Radiation Therapy Planning (RTP) system and its associated software; composed of several modules including the core Pinnacle3 functionality... on . The reason FDA records is "Incorrect Absolute Marking Coordinates when used with PQ-Series scanners due to a software malfunction". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1212-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- The Pinnacle3 Radiation Therapy Planning (RTP) system and its associated software; composed of several modules including the core Pinnacle3 functionality, Syntegra (TM), P3IMRT, P3MD and AcQSim3 (TM). The Pinnacle3 RTP system is composed of a Sun UNIX workstation (or UNIX-complaint computer) running the Solaris operating system and software, and a UNIX terminal emulation package on a personal computer, which allows the user to enter patient data into the system, use that data to construct a plan for radiation therapy, and evaluate the plan- Philips Medical Systems-Model #9200-0652A-ENG, 9200-0700A-ENG, 9200-0700B-ENG, 9200-0700C-ENG (with software versions 7.6C, 8.0d and 8.0h)
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Incorrect Absolute Marking Coordinates when used with PQ-Series scanners due to a software malfunction
- Root cause tag
- software
- FDA root cause
- Other
- Action
- All customers will receive an August 21, 2007, Customer Advisory Notification letter notifying them of the issue and the steps required to avoid the issue. Return receipts will be tracked. This is a short-term fix, a long term software fix will be issued shortly by the firm.
- Quantity in commerce
- 1,407
- Distribution
- Worldwide Distribution; USA including states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, and WI and Country of CANADA
- Product codes and lots
- Pinnacle3 Version 7.6c Software Kit Model #9200-0652A-ENG;Pinnacle3 Version 8.0d Software Kit Model #9200-0700A-ENG;Pinnacle3 Version 8.0d p 1 Software Kit Model #9200-0700B-ENG;Pinnacle3 Version 8.0h Software Kit Model #9200-0700C-ENG