The product is an isolation transformer, an accessory to Computed Tomography Scanner. It consists of...
Philips Medical Systems (Cleveland) Inc began a recall of The product is an isolation transformer, an accessory to Computed Tomography Scanner. It consists of electronic circuitry enclosed in a metal case. According... on . The reason FDA records is "The TEAL Power Conditioning units may fail and result in the unit emitting fire and/or smoke. Note: the TEAL Power Conditioning Unit has the potential to fail when the scanner that is attached to the TEAL is powered and...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1234-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- The product is an isolation transformer, an accessory to Computed Tomography Scanner. It consists of electronic circuitry enclosed in a metal case. According to Philips Medical Systems, the metal case is manufactured from self-extinguishing materials. The unit is labeled in part, "Power Conditioner *** Part Number 4450028-XX *** Frequency: *** S/N: *** Rev: *** MFG date: *** INPUT RATINGS *** Model: MCT 100/480 ISOTRAN PLUS *** OUTPUT RATINGS *** MEDICAL POWER SUPPLY UL *** TEAL Electronics Corporation, San Diego, California ***".
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The TEAL Power Conditioning units may fail and result in the unit emitting fire and/or smoke. Note: the TEAL Power Conditioning Unit has the potential to fail when the scanner that is attached to the TEAL is powered and in use as well as when not in use.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On 9/04/2007, Philips Medical Systems (PMS) sent a letter via courier to all affected customers. The letter notified customers about the failures of the TEAL power conditioner and the possible emisson of fire and/or smoke from the units. In addition, the letter notes that PMS implemented a Field Change Order in May, 2007 that directs onsite repair of all affected units by PMS service technicians.
- Quantity in commerce
- 491
- Distribution
- Worldwide, including, USA, Canada, China, Italy, Russia, Brazil, Belgium, Germany, Austria, Australia, Netherlands, France, Korea, Denmark, Israel, England, Argentina, Mexico, Spain, Yemen, Greece, Egypt, Japan, UAE, & Chile.
- Product codes and lots
- Part Number: 4450028-01; Serial Numbers: 73343, 74482, 74481, 74480, 74477, 74478, 74479, 74639, 74637, 74640, 74638, 74641, 74758, 79281, 79279, 79282, 79278, 79280, 79348, 79347, 79345, 79346, 79512, 79513, 79514, 79511, 79515, 79582, 79699, 79698, 79696, 79697, 79700, 79778, 79774, 79776, 79773, 79775, 79772, 79777, 79898, 79899, 79893, 79894, 79892, 79897, 79896, 79891, 79895, 79976, 79980, 79978, 79975, 79977, 79979, 79974, 80060, 80132, 80125, 80127, 80123, 80124, 80131, 80161, 80128, 80130, 80129, 80126, 80571, 80570, 80574, 80579, 80577, 80572, 80573, 80578, 80575, 80576, 80729, 80730, 80894, 80895, 80896, 80934, 80932, 80897, 80931, 80933, 80935, 80848, 81007, 81004, 81008, 81005, 81010, 81009, 81006, 81090, 81091, 81095, 81093, 81089, 81092, 91094, 746390. Part Number 4450028-02; Serial Numbers: 80583, 81116, 81121, 81117, 81115, 81119, 81128, 81118, 81120, 81126, 81122, 81123, 81125, 81127, 81124, 81129, 81168, 81165, 81169, 81167, 81166, 81170, 81179, 81177, 81171, 81175, 81176, 81351, 81354, 81347, 81348, 81349, 81353, 81350, 81346, 81355, 81352, 81173, 81174, 81172, 81178, 81446, 81454, 81458, 81459, 81455, 81448, 81445, 81449, 81447, 81456, 81457, 81453, 81451, 81452, 81450, 81826, 81830, 81827, 81829, 81828, 81851, 81853, 81852, 81854, 81855, 82053, 82054, 82051, 82052, 82050, 82114, 82112, 82113, 82116, 82119, 82115, 82117, 82120, 82121, 82118, 82154, 82156, 82155, 82157, 82153, 82267, 82266, 82273, 82268, 82265, 82269, 82274, 82271, 82272, 82270, 82387, 82386, 82385, 82383, 82384, 82472, 82473, 82476, 82474, 82475, 82519, 82521, 82518, 82517, 82520, 82604, 82600, 82598, 82603, 82602, 82599, 82607, 82605, 82606, 82601, 82705, 82708, 82707, 82704, 82706, 82798, 82797, 82801, 82799, 82800, 82928, 82927, 82926, 82924, 82925, 83010, 83011, 83012, 83014, 83013, 83109, 83108, 83107, 83106, 83115, 83112, 83110, 83111, 83113, 83114, 83156, 83157, 83158, 83159, 83160, 83155, 83161, 83164, 83162, 83276, 83269, 83272, 83273, 83163, 83271, 83274, 83275, 83277, 83270, 83278, 83381, 83378, 83380, 83375, 83377, 83373, 83374, 83379, 83382, 83376, 73546, 83543, 83548, 83543, 83547, 83544, 83549, 83550, 83551, 83552, 83675, 83676, 83673, 83678, 83681, 83679, 83677, 83674, 83672, 83680, 83748, 83746, 83745, 83744, 83747, 83882, 83878, 83875, 83876, 83877, 83880, 83883, 83884, 83879, 83881, 83881, 83884, 83879, 83880, 83883, 84010, 84014, 84009, 84013, 84008, 84011, 84012, 84016, 84007, 84015, 84065, 84064, 84066, 84072, 84071, 84070, 84069, 84067, 84068, 84073, 84209, 84211, 84212, 84213, 84207, 84210, 84208, 84215, 84216, 84236, 84239, 84233, 84237, 84231, 84235, 84238, 84240, 84232, 84234, 83672, 84519, 84517, 84514, 84516, 84515, 84513, 83877, 83878, 84518, 84520, 84523, 84525, 84526, 84521, 84524, 84522, 84527, 84597, 84598, 84595, 84599, 84596, 84600, 84849, 84854, 84853, 84852, 84850, 84848, 84855, 84851, 84857, 84856, 85014, 85007, 85012, 85006, 85010, 85011, 85009, 85015, 85005, 85016, 85008, 85013, 85017, 85004, 85003, 83678, 85148, 85151, 85149, 85154, 85156, 85152, 85150, 85155, 85157, 85253, 85256, 85258, 85259, 85257, 85254, 85252, 85255, 85153, 85262, 85261, 85260, 85263, 85265, 85264, 85266, 85281, 85282, 85283, 85284, 85286, 85289, 85290, 85287, 85285, 85288, 85320, 85318, 85314, 85313, 85317, 85321, 85316, 85322, 85323, 85325, 85327, 85319, 85333, 85324, 85326, 85332, 85347, 85330, 85315, 85349, 85348, 85336, 85335, 85334, 85337, 85351, 85340, 85353, 85350, 85355, 85339, 85356, 85374, 85352, 85354, 85372, 85371, 85373, 85379, 85378, 85377, 85376, 85375, 85380, 85790, 85791, 85792, 85786, 85785, 85793, 85788, 85784, 85787, 85789.