Z-0084-2008
Foundation Hip System Bipolar Modular, Size 42mm OD; Cat. #412-01-042; Sterile; assembly required by...
Encore Medical, LP / initiated 2007-08-02 / Terminated / root cause: packaging
Encore Medical, Lp began a recall of Foundation Hip System Bipolar Modular, Size 42mm OD; Cat. #412-01-042; Sterile; assembly required by physician prior to implant; manufactured by Encore... on . The reason FDA records is "Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0084-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- Foundation Hip System Bipolar Modular, Size 42mm OD; Cat. #412-01-042; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.
- Product code
- KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
- Reason for recall
- Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.
- Quantity in commerce
- 382 units.
- Distribution
- Worldwide, including USA, Greece, Japan, and Saudi Arabia.
- Product codes and lots
- Lot #: 644021, 648331, 672341, 675901, 678461, 678471, 678481, 678491, 678501, 678511, 678521, 693251, 694501, 697771, 732831, 733871, 736411, 736741, 738071, 738081, 738411, 738421, 770041, 774781, 774791, 777391, 779571, 779581, 783901, 783911, 788351, 794371, 794381, 811151, 811161, 835501, 835511, 836161, 844591, 845951, 848661, 854461, 854471, 854481, 854491, 857681, 874941, 874951, 882561, 882571, 882581, 893551, 893561, 893571, 903121, 903131, 933751, 944241, 944251, 949391, 949401, 964171, 964181, 964191 and 964201.