Z-0085-2008
Foundation Hip System Bipolar Modular, Size 43mm OD; Cat. #412-01-043; Sterile; assembly required by...
Encore Medical, LP / initiated 2007-08-02 / Terminated / root cause: packaging
Encore Medical, Lp began a recall of Foundation Hip System Bipolar Modular, Size 43mm OD; Cat. #412-01-043; Sterile; assembly required by physician prior to implant; manufactured by Encore... on . The reason FDA records is "Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0085-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- Foundation Hip System Bipolar Modular, Size 43mm OD; Cat. #412-01-043; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.
- Product code
- KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
- Reason for recall
- Device mislabeled. Hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.
- Quantity in commerce
- 479 units.
- Distribution
- Worldwide, including USA, Greece, Japan, and Saudi Arabia.
- Product codes and lots
- Lot #: 679631, 678531, 678541, 678551, 678561, 678571, 691661, 692771, 692761, 694351, 694401, 694411, 694421, 698261, 695131, 695911, 697781, 730131, 731451, 731461, 731471, 205452, 736421, 770051, 770061, 770071, 770181, 780151, 783921, 788361, 792671, 794391, 811171, 811181, 811191, 811201, 811291, 832281, 832291, 835521, 842301, 842311, 848671, 854501, 854511, 854981, 874961, 874971, 874981, 882591, 882601, 893581, 893591, 893601, 893611, 903141, 903151, 905561, 925731, 925741, 925751, 927231, 932711, 932721, 933761, 933771, 933781, 944261, 944271, 949411, 949421, 964211, 964221, 964231 and 964241.