Foundation Hip System Bipolar Modular, Size 46mm OD; Cat. #412-01-046; Sterile; assembly required by...

Encore Medical, LP / initiated 2007-08-02 / Terminated / root cause: packaging

Encore Medical, Lp began a recall of Foundation Hip System Bipolar Modular, Size 46mm OD; Cat. #412-01-046; Sterile; assembly required by physician prior to implant; manufactured by Encore... on . The reason FDA records is "Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0088-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
Foundation Hip System Bipolar Modular, Size 46mm OD; Cat. #412-01-046; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.
Product code
KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
Reason for recall
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.
Quantity in commerce
605 units.
Distribution
Worldwide, including USA, Greece, Japan, and Saudi Arabia.
Product codes and lots
Lot #: 644031, 672331, 677061, 677981, 677991, 678691, 678701, 678711, 678721, 679651, 679661, 697671, 691281, 691291, 693261, 699171, 730241, 735791, 736491, 736501, 736511, 736771, 737811, 738441, 738451, 738461, 770101, 770111, 770221, 773281, 777421, 778431, 782551, 788441, 788451, 810911, 811211, 813501, 830751, 830761, 830911, 830921, 835531, 835541, 842361, 842371, 847411, 848681, 848691, 853541, 857731, 857741, 857751, 875571, 875581, 878371, 878381, 878391, 885381, 885401, 892331, 894241, 894251, 905611, 907121, 907131, 907141, 907211, 925801, 925811, 927241, 927251, 927261, 932761, 949471, 949491, 949851, 951101, 951111, 951121, 958941, 966131, 966141 and 982721.