Z-0096-2008
Foundation Hip System Bipolar Modular, Size 54mm OD; Cat. #412-01-054; Sterile; assembly required by...
Encore Medical, LP / initiated 2007-08-02 / Terminated / root cause: packaging
Encore Medical, Lp began a recall of Foundation Hip System Bipolar Modular, Size 54mm OD; Cat. #412-01-054; Sterile; assembly required by physician prior to implant; manufactured by Encore... on . The reason FDA records is "Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0096-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- Foundation Hip System Bipolar Modular, Size 54mm OD; Cat. #412-01-054; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.
- Product code
- KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
- Reason for recall
- Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.
- Quantity in commerce
- 137 units.
- Distribution
- Worldwide, including USA, Greece, Japan, and Saudi Arabia.
- Product codes and lots
- Lot #: 608031, 648321, 675931, 678951, 678961, 678971, 695091, 695851, 780401, 788581, 788591, 788601, 794461, 831021, 840371, 844651, 848711, 855031, 857811, 875551, 894361, 894371, 907281, 933941, 949601, 964501 and 964511.