Z-0097-2008

Foundation Hip System Bipolar Modular, Size 56mm OD; Cat. #412-01-056; Sterile; assembly required by...

Encore Medical, LP / initiated 2007-08-02 / Terminated / root cause: packaging

Encore Medical, Lp began a recall of Foundation Hip System Bipolar Modular, Size 56mm OD; Cat. #412-01-056; Sterile; assembly required by physician prior to implant; manufactured by Encore... on . The reason FDA records is "Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0097-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
Foundation Hip System Bipolar Modular, Size 56mm OD; Cat. #412-01-056; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.
Product code
KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
Reason for recall
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.
Quantity in commerce
86 units.
Distribution
Worldwide, including USA, Greece, Japan, and Saudi Arabia.
Product codes and lots
Lot #: 627751, 629111, 678981, 678991, 679001, 690231, 210192, 695841, 855041, 875561, 882671, 894381, 905671, 9072291, 939741, 949611 and 964521.