Z-0098-2008

Foundation Hip System Bipolar Modular, Size 58mm OD; Cat. #412-01-058; Sterile; assembly required by...

Encore Medical, LP / initiated 2007-08-02 / Terminated / root cause: packaging

Encore Medical, Lp began a recall of Foundation Hip System Bipolar Modular, Size 58mm OD; Cat. #412-01-058; Sterile; assembly required by physician prior to implant; manufactured by Encore... on . The reason FDA records is "Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0098-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
Foundation Hip System Bipolar Modular, Size 58mm OD; Cat. #412-01-058; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.
Product code
KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
Reason for recall
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.
Distribution
Worldwide, including USA, Greece, Japan, and Saudi Arabia.
Product codes and lots
Lot #: 546711, 546721, 546731, 679011, 679021, 379031, 679041, 679051, 695071, 695871 and 989981.