Z-0099-2008

Foundation Hip System Bipolar Modular, Size 60mm OD; Cat. #412-01-060; Sterile; assembly required by...

Encore Medical, LP / initiated 2007-08-02 / Terminated / root cause: packaging

Encore Medical, Lp began a recall of Foundation Hip System Bipolar Modular, Size 60mm OD; Cat. #412-01-060; Sterile; assembly required by physician prior to implant; manufactured by Encore... on . The reason FDA records is "Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0099-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
Foundation Hip System Bipolar Modular, Size 60mm OD; Cat. #412-01-060; Sterile; assembly required by physician prior to implant; manufactured by Encore Medical, L.P., Austin, TX 78758.
Product code
KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
Reason for recall
Device mislabeled; hip prosthesis components validated for 5 years were labeled as having a 6 year expiration date.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Firm notified consignees of recall via phone, e-mail and overnight letters beginning on 08/02/07. Consignees asked to return devices for a properly labeled replacement.
Quantity in commerce
13 units.
Distribution
Worldwide, including USA, Greece, Japan, and Saudi Arabia.
Product codes and lots
Lot #: 677621, 679061, 679071, 679081, 679091, 633841, 695881, 899041 and 989991.