Z-0134-2008

Smiths Medical CADD-Sentry Pro Medication Safety Software - electronic pump protocol programming software...

Abbott Medical / initiated 2007-08-14 / Terminated / root cause: software

Smiths Medical MD, Inc. began a recall of Smiths Medical CADD-Sentry Pro Medication Safety Software - electronic pump protocol programming software - Administrator REF 21-6275-51 RX only, Smiths... on . The reason FDA records is "Incorrect dosage possible: A software anomaly in the CADD-Sentry Pro Administrator software (version B) where the values of a weight based protocol go to zero or round up to the next higher hundredths place, even if...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0134-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Smiths Medical CADD-Sentry Pro Medication Safety Software - electronic pump protocol programming software - Administrator REF 21-6275-51 RX only, Smiths Medical MD, Inc. St Paul, Minnesota 55112 USA
Product code
FRN Pump, Infusion
Reason for recall
Incorrect dosage possible: A software anomaly in the CADD-Sentry Pro Administrator software (version B) where the values of a weight based protocol go to zero or round up to the next higher hundredths place, even if Administrator displayed the values correctly at data entry.
Root cause tag
software
FDA root cause
Software design
Action
A Smiths Medical MD employee visited the consignees between August 14 and 31, 2007, to verbally notify them of the recall, collect the recall CD-ROM containing the affected software and replace with the updated software.
Quantity in commerce
3 CD-ROMs
Distribution
UT, MD, CO
Product codes and lots
Catalog Number: 21-6275-51, Software Version B.