Z-0101-2008

Steris QFC1728 and QFC1729 Quick Connects;(used to connect gastrointestinal endoscopes to processing trays)...

STERIS Corporation / initiated 2007-09-12 / Terminated / root cause: design

Steris Corporation Hopkins Facility began a recall of Steris QFC1728 and QFC1729 Quick Connects;(used to connect gastrointestinal endoscopes to processing trays) The malfunctioning piece of the quick connect... on . The reason FDA records is "Loose or missing adaptor: The center insert of the adaptor #6 restrictor could become loose and/or come out of the adaptor's housing. The absence of adaptor #6's center insert may affect the sterilization process of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0101-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
Steris QFC1728 and QFC1729 Quick Connects;(used to connect gastrointestinal endoscopes to processing trays) The malfunctioning piece of the quick connect component is adaptor #6, STERIS part number 201158. Adaptor #6 attaches to the suction port of the endoscope.
Product code
MED Sterilant, Medical Devices
Reason for recall
Loose or missing adaptor: The center insert of the adaptor #6 restrictor could become loose and/or come out of the adaptor's housing. The absence of adaptor #6's center insert may affect the sterilization process of the attached endoscope.
Root cause tag
design
FDA root cause
Component design/selection
Action
STERIS has contacted affected direct accounts by phone to notify direct accounts of the recall. On 9/28/2007, STERIS will mail a recall notification to all affected direct accounts via certified mail. The notification informs direct accounts that STERIS representatives will be visiting direct accounts to retrieve affected products and provide replacements. Where direct accounts further distribute, STERIS representatives will inform the direct accounts to contact their sub-accounts and facilitate exchange of affected product for replacement product. Replacement product will be available beginning November 1, 2007.
Quantity in commerce
391 Quick Connects
Distribution
Worldwide: United States, Australia, Bahrain, Canada, Germany, Great Britain, Hungary, Italy, Qatar, and Spain
Product codes and lots
C041782, C041793, 3072287, C041514, C041512, C041508, C041513, C041515, 3072303, 3290749, 3338126, C041781, C041794, 3072246, C041506, C041509, C041510, C041511, 3072261, 3290723, 3318292, 3338100, 3352374, 3542412, 3643871, 3546157, 3656121, 3893160, 3890027, 3748456, 3930062, 3978731, 3342342, 3352390, 3575941, 3643897, 3668357, 3895661, 3900198