Z-0204-2009
Terumo Advanced Perfusion System 1 base; Model 801764. NOTE: This product is not distributed in the United...
Terumo Cardiovascular Systems Corporation / initiated 2008-07-18 / Terminated
Terumo Cardiovascular Systems Corporation began a recall of Terumo Advanced Perfusion System 1 base; Model 801764. NOTE: This product is not distributed in the United States. For use for up to 6 hours in the... on . The reason FDA records is "The units may exhibit blank or distorted displays, loss of local control and/or may have local control knobs that are difficult to use". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0204-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1 base; Model 801764. NOTE: This product is not distributed in the United States. For use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- The units may exhibit blank or distorted displays, loss of local control and/or may have local control knobs that are difficult to use.
- FDA root cause
- Other
- Action
- Consignees were notified via letter dated 7/18/08 and instructed to use the central control module should the problem occur. Contact Terumo Cardiovascular Systems Corp at 1-800-521-2818 for assistance.
- Quantity in commerce
- 300
- Distribution
- Nationwide, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.
- Product codes and lots
- Serial numbers 0006 through 0066, 0100 through 0322 and 1001 through 1017.