Z-0065-2008
Bio-Rad Laboratories-MONOLISA Anti-HBc IGM EIA kit, an in-vitro diagnostic test kit, Catalog # 26174...
Bio-Rad Laboratories Inc. / initiated 2007-09-05 / Terminated / root cause: labeling and UDI
Bio-Rad Laboratories began a recall of Bio-Rad Laboratories-MONOLISA Anti-HBc IGM EIA kit, an in-vitro diagnostic test kit, Catalog # 26174, manufactured by Bio-Rad Laboratories, Redmond, WA on . The reason FDA records is "Incorrect Instruction: the package insert states to add 100L of conjugate; it should state to add 100uL". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0065-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bio-Rad Laboratories Inc.
- Product
- Bio-Rad Laboratories-MONOLISA Anti-HBc IGM EIA kit, an in-vitro diagnostic test kit, Catalog # 26174, manufactured by Bio-Rad Laboratories, Redmond, WA.
- Product code
- LOM Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)
- Reason for recall
- Incorrect Instruction: the package insert states to add 100L of conjugate; it should state to add 100uL.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- On 9/5/07 a letter titled "IMPORTANT PRODUCT INFORMATION" was sent to customers advising them of the issue and requesting customers make a note of the correction in their laboratory procedures and/or any existing copies of the package insert. On 9/19/07 the firm sent a second letter (dated 9/18/07) via DHL titled "URGENT PRODUCT CORRECTION" and supplied a corrected package insert to replace the original insert sent.
- Quantity in commerce
- 46
- Distribution
- Nationwide: Kits were distributed to 27 hospitals, clinics and testing labs throughout the US.
- Product codes and lots
- Lot 131BHH, Catalog # 26174