GE LOGIQ 3 Expert ultrasound scanner with software versions 4.0.5, 4.1.1 and 4.1.2. GE Healthcare P.O. Box...
GE Healthcare Clinical Systems began a recall of GE LOGIQ 3 Expert ultrasound scanner with software versions 4.0.5, 4.1.1 and 4.1.2. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201 on . The reason FDA records is "Inaccurate guidelines: The BE9C is an optional, small radius Bi-planar convex probe intended for use with the LOGIQ 3 Expert system in Urology application. An attempt by the user to activate the biopsy guidelines while...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1155-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare Clinical Systems
- Product
- GE LOGIQ 3 Expert ultrasound scanner with software versions 4.0.5, 4.1.1 and 4.1.2. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Inaccurate guidelines: The BE9C is an optional, small radius Bi-planar convex probe intended for use with the LOGIQ 3 Expert system in Urology application. An attempt by the user to activate the biopsy guidelines while this probe is in use will cause incorrect guidelines to be displayed on the image. The incorrect guidelines are recognized by their orientation that is left/right reversed from t
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- GE sent consignees a "GE Urgent Medical Device Correction" letter dated 8/14/07. The letter informed the customers about the problem, offer a short-term workaround and informed them that a representative will schedule a visit to upgrade the software.
- Quantity in commerce
- 62
- Distribution
- Worldwide Distribution - USA (NY & AL), Korea, China, India, Brazil, Mexico, Germany, Greece, Italy, Poland, Russia, Saudi Arabia, Spain, and Czech Republic.
- Product codes and lots
- Systems with software version 4.0.5, 4.1.1 and 4.1.2. Serial numbers: 26022WS4, 26023WS2, 26346WS7, 2641WS8, 31001WS1, 34381WS4, 36460WS4, 36461WS2, 43734WS3, 43735WS0, 43736WS8, 43737WS6, 43994WS3, 43995WS0, 43998WS4, 44000WS8, 44001WS6, 44002WS4, 44003WS2, 44004WS0, 44005WS7, 44011WS5, 44012WS3, 44013WS1, 44015WS6, 44016WS4, 44018WS0, 44019WS8, 44020WS6, 44100WS6, 44153WS5, 44157WS6, 44159WS2, 44270WS7, 44273WS1, 44274WS9, 44275WS6, 44276WS4, 44277WS2, 44532WS0, 44533WS8, 44817WS5, 44819WS1, 44820WS9, 44821WS7, 44822WS5, 44825WS8, 44826WS6, 44908WS2, 44909WS0, 44912WS4, 44914WS0, 44915WS7, 44989WS2, 44990WS0, 44991WS8, 44992WS6, 44994WS2, 45088WS2, 45089WS0, 5122WS7