Z-0319-2008

Biomet Optigun Cement Gun, Ratchet, non-sterile; REF 419500 and REF 4195 (in Europe), Biomet, Inc., Warsaw, IN

Zimmer GmbH / initiated 2007-09-21 / Terminated / root cause: design

Biomet, Inc. began a recall of Biomet Optigun Cement Gun, Ratchet, non-sterile; REF 419500 and REF 4195 (in Europe), Biomet, Inc., Warsaw, IN on . The reason FDA records is "Loose Screw Component: A screw may come loose and fall from the device during use, and result in implantation with the bone cement". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0319-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet Optigun Cement Gun, Ratchet, non-sterile; REF 419500 and REF 4195 (in Europe), Biomet, Inc., Warsaw, IN
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Loose Screw Component: A screw may come loose and fall from the device during use, and result in implantation with the bone cement.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were notified via Urgent Medical Device Correction letter dated 9/21/07 to assure the screw components are fully seated and secure, to not use any guns with loose or missing screws, and to contact Biomet for instructions to return all instruments for evaluation and refurbishment.
Quantity in commerce
6,590 worldwide (2,301 to U.S.) for both recalls Z-0318-0319-2008
Distribution
Worldwide: USA, Canada and Japan.
Product codes and lots
All units including lots 429900, 642100 and 647700.