Z-0319-2008
Biomet Optigun Cement Gun, Ratchet, non-sterile; REF 419500 and REF 4195 (in Europe), Biomet, Inc., Warsaw, IN
Zimmer GmbH / initiated 2007-09-21 / Terminated / root cause: design
Biomet, Inc. began a recall of Biomet Optigun Cement Gun, Ratchet, non-sterile; REF 419500 and REF 4195 (in Europe), Biomet, Inc., Warsaw, IN on . The reason FDA records is "Loose Screw Component: A screw may come loose and fall from the device during use, and result in implantation with the bone cement". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0319-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Biomet Optigun Cement Gun, Ratchet, non-sterile; REF 419500 and REF 4195 (in Europe), Biomet, Inc., Warsaw, IN
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Loose Screw Component: A screw may come loose and fall from the device during use, and result in implantation with the bone cement.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were notified via Urgent Medical Device Correction letter dated 9/21/07 to assure the screw components are fully seated and secure, to not use any guns with loose or missing screws, and to contact Biomet for instructions to return all instruments for evaluation and refurbishment.
- Quantity in commerce
- 6,590 worldwide (2,301 to U.S.) for both recalls Z-0318-0319-2008
- Distribution
- Worldwide: USA, Canada and Japan.
- Product codes and lots
- All units including lots 429900, 642100 and 647700.