Clearview hCG Combo with Reference Line, Wampole Laboratories, ABI List No/Product No. 6028KCV; 30...
Applied Biotech Inc began a recall of Clearview hCG Combo with Reference Line, Wampole Laboratories, ABI List No/Product No. 6028KCV; 30 Individually pouched test cassettes with disposable... on . The reason FDA records is "Sensitivity: Kits contain test devices which may not meet the label claims for sensitivity, therefore use of such devices may produce false negative results with patient samples at or near the limit of detection of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0244-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Applied Biotech Inc
- Product
- Clearview hCG Combo with Reference Line, Wampole Laboratories, ABI List No/Product No. 6028KCV; 30 Individually pouched test cassettes with disposable pipettes, 1 Directional Insert; Inverness Medical Professional Diagnostics Group; Marketed by Wampole Laboratories, 2 Research Way, Princeton NJ 08540 USA
- Product code
- JHI Visual, Pregnancy Hcg, Prescription Use
- Reason for recall
- Sensitivity: Kits contain test devices which may not meet the label claims for sensitivity, therefore use of such devices may produce false negative results with patient samples at or near the limit of detection of the product.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- A Product Recall Notification dated 9/28/2006, was provided via fax, email or direct mail to all primary consignees. Customers were instructed to block all stock of affected lots in their warehouses, contact their primary consignees and advise them to discontinue use immediately and dispose of any unused materials; using a fax back form to communicate to the recalling firm the quantity blocked and disposed of so that arrangements could be made for providing replacement product
- Quantity in commerce
- 400 Kits (1,200 tests)
- Distribution
- Worldwide: USA and Canada
- Product codes and lots
- Model Number: 6028KCV, ABI Lot/Unit Code: 88982