Z-0103-2008
Boston Scientific mach 1 Guide Catheter, 8F, FCL 3.5 SH Femoral Curve Left 3.5, REF 34358-268, UPN...
Boston Scientific Corporation / initiated 2007-09-11 / Terminated / root cause: manufacturing process
Boston Scientific began a recall of Boston Scientific mach 1 Guide Catheter, 8F, FCL 3.5 SH Femoral Curve Left 3.5, REF 34358-268, UPN H749343582680, Sterile EO Made in Mexico Avenida Norske... on . The reason FDA records is "Missing holes: Catheters are missing the catheter's side holes". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0103-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific mach 1 Guide Catheter, 8F, FCL 3.5 SH Femoral Curve Left 3.5, REF 34358-268, UPN H749343582680, Sterile EO Made in Mexico Avenida Norske Edicficio G1, Local B y H Parque Industrial La Mesa, Fracc. Rubio La Mesa, Tijuana, B.C. Mexico C.P. 22550
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- Missing holes: Catheters are missing the catheter's side holes.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A voice mail was left to affected members of the domestic sales force on September 11, 2007 and an interoffice memo with the recall implementation instructions was also sent to the Sales Force. An Urgent Voluntary Medical Device Recall letter dated September 11, 2007, was sent to affected consignees describing the problem and products affected. The letter included the customer steps for recall, a Reply Verification Tracking Form and a return shipping label.
- Quantity in commerce
- 6
- Distribution
- Worldwide: NY, PA, and Japan
- Product codes and lots
- Lot/Batch # 572191