Z-0043-2008

ESA LeadCare II Blood Test Kit, 6 test Kit containing Level 1 and Level 2 Control Vials with lot number...

Magellan Diagnostics, Inc. / initiated 2007-09-20 / Terminated / root cause: manufacturing process

Esa Biosciences, Inc. began a recall of ESA LeadCare II Blood Test Kit, 6 test Kit containing Level 1 and Level 2 Control Vials with lot number: 0704A Catalog Number: 70-7142 on . The reason FDA records is "Level 1 and Level 2 Control values were incorrectly assigned". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0043-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Magellan Diagnostics, Inc.
Product
ESA LeadCare II Blood Test Kit, 6 test Kit containing Level 1 and Level 2 Control Vials with lot number: 0704A Catalog Number: 70-7142
Product code
DOF Lead, Atomic Absorption
Reason for recall
Level 1 and Level 2 Control values were incorrectly assigned.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
ESA notifed Distributors and End Users by telephone and letter on 9/20/07. Distributors and End Users will be informed of the Level 1 and 2 Control value assignment change. End Users will be sent a new Level 1 and 2 Control Vial label for each vial they originally received. Distributors will be asked to return all test kits in their inventory to ESA Biosciences.
Quantity in commerce
2 kits
Distribution
Nationwide and United Kingdom.
Product codes and lots
Lot number: 0704A