Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 80...
Allez Spine, LLC began a recall of Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 80, Catalog Number LG-PR0880, Allez Spine, LLC... on . The reason FDA records is "Screw Shank Failure; separation of the screw shank portion from the pedicle head portion of the Size 8 Pedicle Screw, when torquing down the single piece locking nut during implantation". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0191-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Allez Spine, LLC
- Product
- Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 80, Catalog Number LG-PR0880, Allez Spine, LLC., Irvine, CA
- Product code
- MNI Orthosis, Spinal Pedicle Fixation
- Reason for recall
- Screw Shank Failure; separation of the screw shank portion from the pedicle head portion of the Size 8 Pedicle Screw, when torquing down the single piece locking nut during implantation.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On July 30, 2007 Allez Spine notified all its field representatives and distributors via email of a potential for failure of the Size 8 Pedicle Screws when used in combination with the single piece locking nuts and instructed them to return all unused inventory of Size 8 Screws and the two piece Locking Nuts with saddle. This action was to be carried out to the retail level-hospital, clinic, lab, and physicians sold the product. The email also had an attachment of a list of expected frequently asked questions (FAQ) from surgeons and answers to help them respond. A second Notice to all users was sent by the recalling firm on Feb 29, 2008, to request follow-up of implanted units.
- Quantity in commerce
- 14
- Distribution
- Nationwide: including the states of AZ, CA, CO, GA, NV, TX and UT
- Product codes and lots
- Lot Number: 10528