Z-0078-2008

Spectra Optia apheresis system, a blood component separator. Part No. 61000, Gambro BCT Lakewood, CO USA 80215

Terumo BCT, Ltd / initiated 2007-09-18 / Terminated / root cause: software

Gambro BCT, Inc. began a recall of Spectra Optia apheresis system, a blood component separator. Part No. 61000, Gambro BCT Lakewood, CO USA 80215 on . The reason FDA records is "Under specific conditions, apheresis machine could return up to 60mL of air to the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0078-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo BCT, Ltd
Product
Spectra Optia apheresis system, a blood component separator. Part No. 61000, Gambro BCT Lakewood, CO USA 80215.
Product code
LKN Separator, Automated, Blood Cell And Plasma, Therapeutic
Reason for recall
Under specific conditions, apheresis machine could return up to 60mL of air to the patient.
Root cause tag
software
FDA root cause
Software design
Action
All consignees were notified by telephone beginning 09/17/2007. They were given instructions on how to safely use the machine. Letters were also sent on 09/22/2007 restating the information.
Quantity in commerce
15 units
Distribution
US, Canada, Australia, Europe.
Product codes and lots
Part No. 61000, Serial Numbers 1P00100 to 1P00141.