COULTER GEN*S System, Part Number: 6605381, Beckman Coulter, Brea, CA. (Hematology analyzer)
Beckman Coulter Inc began a recall of COULTER GEN*S System, Part Number: 6605381, Beckman Coulter, Brea, CA. (Hematology analyzer) on . The reason FDA records is "Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0315-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- COULTER GEN*S System, Part Number: 6605381, Beckman Coulter, Brea, CA. (Hematology analyzer)
- Product code
- GKL Counter, Cell, Automated (Particle Counter)
- Reason for recall
- Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2)- When manually entering a Patient ID, if a blank space is entered between th
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Product Corrective Action (PCA) letters were mailed on April 18,2007, to all GENS, LH500, LH750 and LH780 accounts. They are informed that a potential misidentification can occur in 2 scenarios. 1. When a positive identifier (Sample ID or CassetteIPosition) is manually edited to a positive identifier that is already in the ToDo list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2. When manually entering a Patient ID, if a blank space is entered between the characters in the Patient ID field, the system will only accept the characters before the space.----Short term fix----- The letter explains the issues and provides the customer work around. (Letters were sent by US mail). A customer response form was provided. The root cause of the problem is still under investigation, software upgrades, when approved will be made to devices.
- Quantity in commerce
- 815 units to the US; 95 units to Canada
- Distribution
- World wide: USA and Canada
- Product codes and lots
- Part Number 6605381. All software versions