Z-0178-2008
Capnostream20 Monitor with Masimo Pulse Oximetry Module - Part Number: CS 08659 - Oridion Medical
Oridion Capnography, Inc. / initiated 2007-07-28 / Terminated / root cause: manufacturing process
Oridion Capnography Inc began a recall of Capnostream20 Monitor with Masimo Pulse Oximetry Module - Part Number: CS 08659 - Oridion Medical on . The reason FDA records is "Handle Cracking - Cracks in the handle will lead to disconnection of the handle from the monitor, causing the monitor to fall". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0178-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Oridion Capnography, Inc.
- Product
- Capnostream20 Monitor with Masimo Pulse Oximetry Module - Part Number: CS 08659 - Oridion Medical
- Product code
- DQA Oximeter
- Reason for recall
- Handle Cracking - Cracks in the handle will lead to disconnection of the handle from the monitor, causing the monitor to fall.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Oridion sent Dear Customer letters dated, August 21, 2007, informing users that a new monitor with the improved handle will be shipped by August 31, 2007.
- Quantity in commerce
- 6
- Distribution
- Nationwide Distribution- including states of AZ, CA, FL, MA, MO, OH, TX, WA, and WI,
- Product codes and lots
- Serial Numbers: B300001003 B300001008 B300001018 B300001043 B300001044 B300001050