Z-1526-2008

GE Datex-Ohmeda Neonatal Enhancement for the Engstrom Carestation, Model # 1505-3272-00, GE Healthcare P.O...

GE Healthcare / initiated 2007-07-02 / Terminated / root cause: manufacturing process

Datex - Ohmeda, Inc began a recall of GE Datex-Ohmeda Neonatal Enhancement for the Engstrom Carestation, Model # 1505-3272-00, GE Healthcare P.O. Box 7550, Madison, WI 53707-7550, USA Intended to... on . The reason FDA records is "Neonatal Flow Sensors supplied for use with the Neonatal Enhancement for the Engstrm Carestation may result in the sensors providing invalid values and alarm messages when exposed to high flow rates". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1526-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Datex-Ohmeda Neonatal Enhancement for the Engstrom Carestation, Model # 1505-3272-00, GE Healthcare P.O. Box 7550, Madison, WI 53707-7550, USA Intended to provide mechanical ventilation for adults and pediatrics weighing 5kg and above having degrees of pulmonary impairment varying from minor to severe.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Neonatal Flow Sensors supplied for use with the Neonatal Enhancement for the Engstrm Carestation may result in the sensors providing invalid values and alarm messages when exposed to high flow rates.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Consignees were sent a Medical Device Correction Notification letter on July 2, 2007. The firm's short term recommendation is that Neonatal Flow Sensors may continue to be used with added viligance around potential patient circuit disconnections and the associated alarms. Ther firm recommends that the MV exp high and Circuit leak alarms be set appropriately for each patient in order to detect a patient disconnect. The long term recommendation was to order the replacement part when available and dispose of the old part when the replacement is received.
Quantity in commerce
394
Distribution
Worldwide Distribution --- USA including state of TX and countries of Austria, Bangladesh, Belgium, Costa Rica, Denmark, Finland, France, Germany, India, Italy, Mexico, Saudi Arabia, South Africa, Sweden, Switzerland, United Kingdom, United Arab Emirates.
Product codes and lots
All serial numbers