Z-0461-2008

Philips Medical Systems, GEMINI TF 16 & 64 slice PET/CT Systems, diagnostic imaging systems for fixed or...

Philips North America LLC / initiated 2007-09-21 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Philips Medical Systems, GEMINI TF 16 & 64 slice PET/CT Systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission... on . The reason FDA records is "Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0461-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Medical Systems, GEMINI TF 16 & 64 slice PET/CT Systems, diagnostic imaging systems for fixed or mobile installations that combine Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems, Philips Medical Systems, Cleveland, OH 44143
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.
Root cause tag
software
FDA root cause
Software design
Action
Philips Medical notified domestic customers via Product Safety Notification dated May 30, 2007. The product safety notification includes a Technical Bulletin notifying customers of the issue and how to avoid it. This information will be used until the software correction becomes available. The software correction will be released in October 2008. Once software is available, the firm will begin correcting the field.
Quantity in commerce
61
Distribution
Worldwide Distribution - USA, Australia, Belgium, Canada, China, Denmark, England, Finland, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Jordan, Korea, Netherlands, Poland, Portugal, Saudi Arabia, Slovenia, Spain, Sweden, Taiwan, Turkey, and Venezuela.
Product codes and lots
Gemini TF   Product Numbers/Manufacturer Numbers: TF 64 slice: 4535 679 94741, TF 16 slice: 4535 679 83931.  Catalog Numbers: TF 64 slice: 882471, TF 16 slice: 882470.   Serial Numbers: 7001-7058, 7062, 7064, and 7066.