AmniSure ROM Test, Fetal Membrane Rupture Test for In-Vitro Diagnostic Use, Model: FMRT-1, AmniSure...
Amnisure International, LLC / initiated 2007-10-03 / Terminated / root cause: design
AmniSure International LLC began a recall of AmniSure ROM Test, Fetal Membrane Rupture Test for In-Vitro Diagnostic Use, Model: FMRT-1, AmniSure International LLC, Cambridge, MA 02138 on . The reason FDA records is "Weak true postive/false negative results: weak true positive or false negatives in ruptured fetal membranes may occur due to a "hook effect"". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0120-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Amnisure International, LLC
- Product
- AmniSure ROM Test, Fetal Membrane Rupture Test for In-Vitro Diagnostic Use, Model: FMRT-1, AmniSure International LLC, Cambridge, MA 02138
- Product code
- JJX Single (Specified) Analyte Controls (Assayed And Unassayed)
- Reason for recall
- Weak true postive/false negative results: weak true positive or false negatives in ruptured fetal membranes may occur due to a "hook effect"
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- AmniSure notified users by Medical Device Correction (Recall) letter dated October 1, 2007 issued via US Mail. The letter provides additional labeling instructions and a checklist for trouble shooting, including a diluation procedure that allows users to determine and solve the problem of the hook effect with this device.
- Quantity in commerce
- 319,976 units
- Distribution
- Worldwide, USA, Australia, Austria, Italy, Korea, Bulgaria, Russia, Spain, Kuwait, Switzerland, Luexemburg, Ireland, Denmark, Netherlands, Jordan, Israel, Chile, UK, Singapore, Basil, New Zealand, Turkey, France, Greece, and Equador
- Product codes and lots
- Lot Number/ Exp. Date A7003, 01/2009. A7002, 11/2008, A7001, 10/2008, A6005, 04/2008, A6004, 01/2008, A6003, 09/2007, and A6002-3, 09/2007