Z-0109-2008
Boston Scientific-Renegade Hi-Flo Microcatheter Kit, UPN Material Number: MOO1183030,Catalog Number: 18-303
Boston Scientific Corporation / initiated 2007-09-10 / Terminated / root cause: packaging
Boston Scientific Corporation began a recall of Boston Scientific-Renegade Hi-Flo Microcatheter Kit, UPN Material Number: MOO1183030,Catalog Number: 18-303 on . The reason FDA records is "Incorrect Guidewire: Product may incorrectly contain a 135 cm length Transend-18 guidewire instead of the kit labeled 165 cm length Transend-18 guidewire". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0109-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific-Renegade Hi-Flo Microcatheter Kit, UPN Material Number: MOO1183030,Catalog Number: 18-303
- Product code
- KRA Catheter, Continuous Flush
- Reason for recall
- Incorrect Guidewire: Product may incorrectly contain a 135 cm length Transend-18 guidewire instead of the kit labeled 165 cm length Transend-18 guidewire.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Boston Scientific issued a Urgent Medical Device Recall Letter dated September 10, 2007 to Domestic accounts via Federal Express. The notification instructed the account to immediately discontinue the use of the product under this recall, check their inventory, remove any affected product , and return it to Boston Scientific. The notification also requested the completion of a Reply Verification Tracking Form
- Quantity in commerce
- 153
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 9218692; 9218628; 9218451