Z-0109-2008

Boston Scientific-Renegade Hi-Flo Microcatheter Kit, UPN Material Number: MOO1183030,Catalog Number: 18-303

Boston Scientific Corporation / initiated 2007-09-10 / Terminated / root cause: packaging

Boston Scientific Corporation began a recall of Boston Scientific-Renegade Hi-Flo Microcatheter Kit, UPN Material Number: MOO1183030,Catalog Number: 18-303 on . The reason FDA records is "Incorrect Guidewire: Product may incorrectly contain a 135 cm length Transend-18 guidewire instead of the kit labeled 165 cm length Transend-18 guidewire". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0109-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific-Renegade Hi-Flo Microcatheter Kit, UPN Material Number: MOO1183030,Catalog Number: 18-303
Product code
KRA Catheter, Continuous Flush
Reason for recall
Incorrect Guidewire: Product may incorrectly contain a 135 cm length Transend-18 guidewire instead of the kit labeled 165 cm length Transend-18 guidewire.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Boston Scientific issued a Urgent Medical Device Recall Letter dated September 10, 2007 to Domestic accounts via Federal Express. The notification instructed the account to immediately discontinue the use of the product under this recall, check their inventory, remove any affected product , and return it to Boston Scientific. The notification also requested the completion of a Reply Verification Tracking Form
Quantity in commerce
153
Distribution
Nationwide
Product codes and lots
Lot Numbers: 9218692; 9218628; 9218451