Z-0046-2008

Microtainer Safety Flow Lancets; Blue Sterile Use once and Discard; 1.9mm Blade Length; 1.0 mm Blade Width...

Becton Dickinson & Co / initiated 2007-09-26 / Terminated / root cause: manufacturing process

Becton Dickinson & Company began a recall of Microtainer Safety Flow Lancets; Blue Sterile Use once and Discard; 1.9mm Blade Length; 1.0 mm Blade Width; Becton Dickinson Vacutainer Systems Becton... on . The reason FDA records is "Potential for the blade and holder assembly to become loose and then separate from the main body of the lancet during the process of disposal". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0046-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
Microtainer Safety Flow Lancets; Blue Sterile Use once and Discard; 1.9mm Blade Length; 1.0 mm Blade Width; Becton Dickinson Vacutainer Systems Becton dickinson and COmpany; Franklin Lakes, NJ 07417
Product code
F
Reason for recall
Potential for the blade and holder assembly to become loose and then separate from the main body of the lancet during the process of disposal.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Recall communication issued on or about 9/28/2007.
Quantity in commerce
4,349,000 units
Distribution
Product distributed worldwide to Australia, Brazil, Canada, China, Colombia, Guatemala, Hong Kong, India, Indonesia, Mexico, Malaysia, New Zealand, Pakistan, Philippines, Singapore, Thailand, and United States.
Product codes and lots
Catalog number 366357; Lot numbers: 2228742; 3022532; 3037536; 3055877; 3062548; 3079392; 3086096; 3135003; 3142723; 3199645; 3225646; 3240751; 3248047; 3268651; 2323030; 2330450.